Product Description

16L / Min Flow Digital Temperature Controlled Recirculating Water Chiller 

 

TMAX-KJ5000is a CE certified Water-cooled Chiller with built in the refrigerating machine that provides a clean, environmentally friendly and reliable source of temperature-controlled fluid for closed loop systems up to 16L/min. It eliminates scale buildup, sewerage cost, and tap water cooling expenses. This small chiller provides high cooling capacity at low cost and has a convenient digital reading and setting display. Temperature range can be adjusted from 5 to 30°C.

SPECIFICATIONS:
Power Supply
AC 100-120V or AC 208-240V can be selected in options bar (50/60Hz).
Current
3.5-5.6A
Refrigerant & Cooling
Capacity
  • Refrigerant: R134a (~300g)
  • Capacity:    722Kcal/h;  2866Btu/h;  840W
Temperature Control
5~30 °C
Temperature Accuracy
± 0.3 °C
Max. Pump Head
10 meters (14psi)
Pipe Connectors
  • Φ 12mm PTC connector
  • 8mm I.D, 12mm O.D. polyurethane tube is included
Water Flow Rate
16 Liters / minute
Water Tank
  • 6 Liters Water Tank made of stainless steel
  • Water chiller should not be connected to a system with the capacity more than the

tank's capacity itself.

  • Use coolant instead of water for longer service life.
  • Do not use deionized water as the cooling agent, DI water may cause the copper pipe

to corrode, mainly pitting, and leak over time.

Water Filter
  • One water filter is included for immediate use.
  • Replace water filter every 3 to 6 months based on water quality and usage. Please

click on pic to order

Protection
Protected from overflow alarm and over temperature
Noise Level
85dBA (standing next to the unit)
75dBA (standing at 1 meter away from the machine)
Overall Dimensions
550 X 280 X 430 mm (L x W x H)
Shipping Weight
100 lbs
Warranty
One year limited warranty with lifetime support
Compliance
  • CE approval
  • MET Certification (UL 1450) is available upon request at extra cost, please contact our

sales representatives for a quote.

Application Notes

Must use Rust Protection coolants to prevent rust and corrosion. The use of tap water is prohibited.

Please connect the chiller water outlet to the equipment water inlet, and the chiller water inlet to

the equipment outlet. Otherwise, the water flow sensor on the equipment may fail to work due

to wrong water flow direction

If you need temperature below 20C, please set Parameter F9 to the desired temperature(10C-20C).

battery equipment